Bently Panel Measurements

Applies to: vb5, vb6, vb7, vb8, vb1000, vb2000, vb3000

QUESTION

How do I read directly from a Keyphasor® output on a Bently panel?

ANSWER

The Commtest instruments support the connection to a Bently panel to read the Keyphasor® signal and the displacement output signals from the installed proximity probes.

In order to make absolutely certain that the connection of the instrument will not cause a voltage spike to the installed system; it is recommended that the measurement be set up on the second channel as a displacement measurement with no power applied.

This is accomplished in Ascent in the parameter setup screen as shown below.

Bently1

The sensor setup is as follows:

bently2

Note that drive current is not enabled. (on vbX000 instruments the Input Coupling is limited to AC only, so the sensor gap cannot be measured)

In the instrument, it is necessary to have CH2 enabled, as shown below:

bently3

A straight-thru male BNC to male BNC cable can then be used for acquiring the data.

For measurements where it is desired to take displacement readings on both channels for an orbit display, both channels would have to be set up for displacement and the appropriate scaling, say 200 mv per mil.

To be absolutely certain that no drive current is present from a previous reading with a default sensor on CH1, it is necessary to take an initial reading without connection to the panel and then a re-take or over-write that data with the proper connection.

For connectivity to the Keyphasor® output, the vbX000 requires a special cable CABK0108 which conditions the signal to match its TTL-style input. The vbX instruments have a different connector, so require cable KEY70258. The vbX instruments have the conditioning electronics built-in, and can be told which type of conditioning to use via the Tach Trigger setup option, as shown below.

bently4

 

FAQ ID: 12010 Last Reviewed: 15 January 2010

 

Back to Frequently Asked Questions